CPSC Recalls

Recalls / 2010-12-14

Extech Instruments Recalls Digital Clamp and Multimeters due to Electrocution Hazard

11067 · Recalled 2010-12-14 · CPSC page

Clamp Meter
Clamp Meter

Description

This recall involves Extech digital clamp meters with model numbers EX612, EX613, EX622 and EX623 and multimeters with model numbers EX540, EX542 and EX570. "Extech" and the model number can be found on the front of the product and the serial number on the rear. Serial numbers that begin with "A" are not included in the recall.

Hazards

When the battery runs low, the meters can fail to give an accurate voltage reading, resulting in the operator falsely believing the electrical power is off or low. This poses an electrocution hazard.

Reported injuries

Extech received one report of a meter displaying an incorrect voltage reading. No injuries have been reported.

Remedy

Consumers should immediately stop using the recalled digital clamp meters and multimeters and contact Extech for a free replacement meter.

Products

NameModelUnits
Digital clamp meters and multimeters5,100 units

Parties

Manufacturer
Extech Instruments
Made in
China
Sold at
Industrial, electrical distributors, wholesalers nationwide, at tool, test equipment retailers, $300.

Consumer contact

For additional information, contact Extech toll-free at (855) 239-8324 between 8:30 a.m. and 5 p.m. ET Monday through Friday or visit the company's website at http://www.extech.com/recall

More images

Multimeters
Multimeters

Source: CPSC recall 11067. Inclusion in this database is not legal or medical advice.